Remote Commands
R&S
®
FSW
1284User Manual 1173.9411.02 ─ 43
Manual operation: See "To - From" on page 694
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:SLISt<sli>:REMove
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:SLISt<sli>:REMove
[SENSe:]CORRection:FRESponse<si>:USER:SLISt<sli>:REMove
Removes the specified touchstone file from the list.
Suffix:
<si>
.
1..n
irrelevant
<sli> 1..n
Index in Touchstone file list
Use [SENSe:]CORRection:FRESponse<si>:USER:
SLISt<sli>:SIZE? to determine the maximum index number.
Example:
SENS:CORR:FRES:USER:SLIS2:REM
Usage: Event
Manual operation: See "Remove File" on page 694
[SENSe:]CORRection:FRESponse<si>:USER:PSTate <State>
Activates or deactivates the preview of the user correction files for all input types.
Note that this function is only available for remote operation. The preview cannot be
switched back on in manual operation.
Suffix:
<si>
.
1..n
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 1
Example:
SENS:CORR:FRES:USER:PST ON
Manual operation: See "Preview" on page 696