Remote Commands
R&S
®
FSW
1297User Manual 1173.9411.02 ─ 43
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation: See "GPIB Terminator" on page 788
SYSTem:DISPlay:LOCK <State>
Defines whether the "Display Update" function remains available in remote operation
or not.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
The function remains available.
ON | 1
The function is not available and the display is not updated dur-
ing remote operation.
*RST: 0
Manual operation: See "Remote Display Update" on page 788
SYSTem:DISPlay:UPDate <State>
This command turns the display during remote operation on and off.
If on, the R&S FSW updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:DISP:UPD ON
Manual operation: See "Remote Display Update" on page 788
SYSTem:ERRor:DISPlay <State>
This command the error display during remote operation on and off.
If activated, the R&S FSW displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:ERR:DISP ON
Manual operation: See "Display Remote Errors" on page 789
SYSTem:IDENtify:FACTory
This command resets the query to *IDN? to its default value.
Configuring the R&S