1298User Manual 1173.9411.02 ─ 43
Manual operation: See "Reset to Factory String" on page 787
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 787
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State> ON
LLO (local lockout)
OFF
GTL (go to local)
*RST: OFF
Example:
SYST:KLOK ON
Activates LLO (remote control)
Manual operation: See "Local" on page 796
SYSTem:LXI:INFO
This command queries the LXI settings.
Return values:
<LXIInfo> <current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current LXI parameters.
• <version>
• <LXIClass>
• <ComputerName>
• <MACAddress>
• <IPAddress>
• <AutoMDIX>
Manual operation: See "Current LXI Configuration" on page 794
SYSTem:LXI:LANReset
This command resets the LAN configuration as required by the LXI standard. The com-
mand also resets the LXI password and instrument description.
Manual operation:
See "LAN Reset" on page 794
Configuring the R&S FSW

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S FSW Series and is the answer not in the manual?

R&S FSW Series Specifications

General IconGeneral
BrandR&S
ModelFSW Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals