Network and Remote Operation
R&S
®
FSW
788User Manual 1173.9411.02 ─ 43
Rohde&Schwarz,FSW,<Unique number>
Remote command:
SYSTem:IDENtify:FACTory on page 1297
Remote Display Update
Defines whether the display of the R&S FSW is updated when changing from manual
operation to remote control.
Turning off the display update function improves performance during remote control.
Note: Usually, this function remains available on the display during remote operation.
However, it can be disabled remotely. In this case, the display is not updated during
remote operation, and cannot be turned on again locally until local operation is
resumed.
Remote command:
SYSTem:DISPlay:UPDate on page 1297
SYSTem:DISPlay:LOCK on page 1297
GPIB Terminator
Changes the GPIB receive terminator.
"LFEOI"
According to the standard, the terminator in ASCII is <LF> and/or
<EOI>.
"EOI"
For binary data transfers (e.g. trace data) from the control computer
to the instrument, the binary code used for <LF> might be included in
the binary data block, and therefore should not be interpreted as a
terminator in this particular case. This can be avoided by using only
the receive terminator EOI.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator on page 1296
*IDN Format
Defines the response format to the remote command *IDN? (see *IDN?
on page 820). This function is intended for re-use of existing control programs
together with the R&S FSW.
"Leg"
Legacy format, as in the R&S FSP/FSU/FSQ family.
"New"
R&S FSW format.
Remote command:
SYSTem:FORMat:IDENt on page 1306
I/O Logging
Activates or deactivates the SCPI error log function. All remote control commands
received by the R&S FSW are recorded in a log file. The files are named according to
the following syntax:
C:\R_S\INSTR\ScpiLogging\ScpiLog.<no.>
where <no.> is a sequential number
A new log file is started each time logging was stopped and is restarted.
Network and Remote Control Settings