Remote Commands
R&S
®
FSW
843User Manual 1173.9411.02 ─ 43
<sb> Sub block in a Multi-standard radio measurement;
MSR ACLR: 1 to 8
Multi-SEM: 1 to 8
for all other measurements: irrelevant
Return values:
<Standards> List of standard files.
Usage: Query only
Manual operation: See "User Standards" on page 169
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:DELete
<Standard>
This command deletes a file containing an ACLR standard.
Suffix:
<n>
.
Window
<m> Marker
<sb> irrelevant
Parameters:
<Standard> String containing the file name of the standard.
Manual operation: See "User Standards" on page 169
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE <Standard>
This command saves the current ACLR measurement configuration as a new ACLR
standard.
The measurement configuration for power measurements consists of weighting filter,
channel bandwidth and spacing, resolution and video bandwidth, detector and sweep
time.
Suffix:
<n>
.
Window
<m> Marker
<sb> irrelevant
Parameters:
<Standard> String containing the file name. The file format is xml.
Manual operation: See "User Standards" on page 169
14.5.3.2 Configuring the Channels
The following commands configure channels for channel power and ACLR measure-
ments.
Configuring and Performing Measurements

Table of Contents