Remote Commands
R&S
®
FSW
998User Manual 1173.9411.02 ─ 43
14.5.14.4 Configuring EMI Limit Lines
The commands required to define limit lines for EMI measurements are described in
Chapter 14.8.4, "Configuring Display and Limit Lines", on page 1209.
14.5.14.5 Controlling LISN
The commands required to control a LISN are described here.
INPut<ip>:LISN:FILTer:HPASs[:STATe].............................................................................998
INPut<ip>:LISN:PHASe.................................................................................................. 998
INPut<ip>:LISN[:TYPE]...................................................................................................999
INPut<ip>:LISN:FILTer:HPASs[:STATe] <State>
This command turns the 150 kHz highpass filter for the ENV216 network on and off.
Suffix:
<ip>
.
1 | 2
For R&S FSW85 models with two RF input connectors:
1: Input 1 (1 mm [RF Input] connector)
2: Input 2 (1.85 mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Example: //Turn on high pass filter
INP:LISN:TYPE ENV216
INP:LISN:FILT:HPAS ON
Manual operation: See "150 kHz Highpass" on page 350
INPut<ip>:LISN:PHASe <Phase>
This command selects one LISN phase to be measured.
Configuring and Performing Measurements