R&S
®
CMW 500 GPRF Applications
GPRF Measurements and Generators
Operating Manual 1202.3986.32 – 03 138
Retrigger All
Like "Retrigger Sweep", however, the trigger is re-armed after each power/frequency
step.
This mode is recommended if for steps with unknown or irregular distances.
7.1.2.4 Measurement Settings
The parameters for the "Power" measurement are accessible from the "Power
Configuration" dialog; see Power Parameters. In the measurement menu, various
softkey with associated hotkeys provide fast access to the most important power
parameters.
The "Power" measurement can be performed with a wide range of IF filters of various
shapes and bandwidths. In addition to Gaussian and bandpass filters, the
measurement provides special filter types for WCDMA and CDMA200 measurements.
7.1.2.5 Power: Results
The basic "Power" measurement result is the power of the RF input signal averaged
over the "Measurement Length". All power results of the sweep can be displayed in a
bar graph or in tables. Moreover, the measurement provides a statistical evaluation at
a particular power step; see Statistical Results.
The statistical evaluation includes the calculation of the "Standard Deviation". This
quantity shows how widely spread the power values are at each power step.