R&S
®
CMW 500 Preparing for Use
Rear Panel Tour
Operating Manual 1202.3986.32 – 03 8
1.1.8.2 LAN Connector
8-pin connector RJ-45 used to connect the R&S CMW 500 to a Local Area Network
(LAN). The LAN connector is equivalent to the connector of the same type labeled
LAN REMOTE on the rear panel of the instrument. Refer to Remote Control in a LAN
and LAN Interface.
1.1.8.3 SENSOR Connector
Lemosa 6-pin socket for NRP-Z27/-Z28 power sensors. An external power sensor can
be used to monitor the power of an external signal or calibrate an RF source; see
External Power Sensor Measurement.
The SENSOR connector controls the power sensor but also provides
and trigger signals.
1.1.8.4 USB Connectors
Single Universal Serial Bus connectors of type A (master
USB), used to connect e.g. a
keyboard, mouse or other pointing devices, a printer or an external storage device
(USB stick, CD-ROM drive etc.). The USB interface can also be used for remote
control of the instrument; see section Remote Control Operation in the "Remote
Control" chapter. All front panel USB connectors comply with standard USB 2.0; refer
to the "Specifications".
USB Connection
The length of passive connecting USB cables should not exceed 1 m. The maximum
current per USB port is 500 mA.
1.1.8.5 AF Connectors
For future extensions.
1.2 Rear Panel Tour
This section gives an overview of the rear panel controls and connectors of the R&S
CMW 500.