R&S
®
CMW 500 WiMAX Applications
WiMAX GUI Reference
Operating Manual 1202.3986.32 – 03 383
Multi-evaluation configuration: trigger settings
Trigger – Trigger Source
Selects the source of the trigger event.
TSRC_POWER_NARR
OWBAND
The measurement is triggered by the RF input power. The
"Trigger Level" setting is valid for this trigger mode; it should
be adjusted to the measured RF signal.
GPIB command:
CONFigure:WIMax:MEASurement<i>:MEValuation:TRIGger:SOURce
Trigger – Trigger Level
Defines the trigger threshold for the "TSRC_POWER_NARROWBAND" trigger relative
to the "Expected Nominal Power". The trigger threshold defines the measured input
signal power where the trigger condition is satisfied and a trigger event is generated.
GPIB command:
CONFigure:WIMax:MEASurement<i>:MEValuation:TRIGger:LEVel
10.2.1.8 WiMAX Multi Evaluation: Results
The results of the WiMAX multi-evaluation measurement are displayed in a single
view.
Table of Results
The measurement diagram shows the most important "General Settings" of the
measurement; refer to WiMAX Multi Evaluation: Parameters. Below, a table contains
quantities to assess the modulation accuracy, frequency error, power, crest factor, and
spectral flatness in the measured burst. Moreover, it provides a statistical evaluation
over a configurable number of successive bursts; see Statistical Results.