R&S
®
CMW 500 WiMAX Applications
WiMAX Command Reference
Operating Manual 1202.3986.32 – 03 396
Characteristics Firmware version V1.00
SCPI: Query only
FETCh:WIMax:MEAS<i>:MEValuation:STATe:ALL?
Queries the main measurement state and the measurement substates. Both
measurement substates are relevant for running measurements only. Use
FETCh:...:MEAS<i>:...:STATe? to query the main measurement state only. Use
INITiate..., STOP..., ABORt... to change the measurement state.
Returned values:
<Main State>, OFF: measurement switched off, no resources allocated, no
results available (when entered after STOP...)
RDY: measurement has been terminated, valid results may be
available
RUN: measurement running (after INITiate..., READ...),
synchronization pending or adjusted, resources active or queued
(see Measurement Substates)
Def. value: OFF
<Sync State>, PEND: waiting for resource allocation, adjustment, hardware
switching ("pending")
ADJ: all necessary adjustments finished, measurement running
("adjusted")
INV: not applicable because <main_state>: OFF or RDY
("invalid")
<Resources State> QUE: measurement without resources, no results available
("queued")
ACT: resources allocated, acquisition of results in progress but
not complete ("active")
INV: not applicable because <main_state>: OFF or RDY
("invalid")
Characteristics Firmware version V1.00
SCPI: Query only
10.3.2.3 Statistical Settings
The following commands specify the scope of the multi-evaluation measurement.
CONFigure:WIMax:MEAS<i>:MEValuation:REPetition <Mode>
Specifies the repetition mode of the measurement; see Statistical Settings. The