R&S
®
CMW 500 Command Reference
Instrument-Control Commands
Operating Manual 1202.3986.32 – 03 125
Example STAT:QUES:PTR 1536
Set bits no. 9 and 10 of the QUEStionable:PTRansition register
(see decimal representation of a bit pattern).
Characteristics Firmware version V1.00
STATus:QUEue[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of SYSTem:ERRor[:NEXT]?
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard.
SCPI Confirmed, query only
Example STAT:QUE?
Query the oldest entry in the error queue. 0,"No error" is
returned if the error queue is empty.
Characteristics Firmware version V1.00
*RST value: –
6.3.4 SYSTem Commands
The SYSTem subsystem contains functions that are not related to instrument
performance. Most of the commands are SCPI-confirmed.
LAN Services
Miscellaneous Instrument Settings
6.3.5 LAN Services
The following commands define the parameters for LAN connection of the R&S CMW
500. .
SYSTem:COMMunicate:NET:ADAPter '<Adapter>'
Selects the network adapter and thus the connection type. At present ,
LAN connections are supported.
'<Adapter>' String parameter, see example below.
Def. value: – (a *RST does not affect the LAN service settings)
Example SYSTem:COMMunicate:NET:ADAPter "Local Area
Connection"
Select a local area (LAN) connection.