CMW 500 GSM Applications
GSM Command Reference
Operating Manual 1202.3986.32 – 03 265
<Status 0> |
<Freq Offset 0>,
...,
<Status 19> |
<Freq Offset 19>
Status:
ON: Enable measurement with the corresponding frequency
offset
OFF: Disable measurement with the corresponding frequency
offset
Def. value, Status 0 to 10: ON
Def. value, Status 11 to 19: OFF
Frequency offset:
Set frequency offset. The status is left unchanged.
Range: 0 MHz to 3 MHz
Def. value, Offset 0 to 10 [MHz]: 0.1, 0.2, 0.25, 0.4, 0.6, 0.8, 1,
1.2, 1.4, 1.6, 1.8
Def. value, Offset 11 to 19 [MHz]: 1.9
Def. unit: MHz
Example See Configuring a Spectrum Measurement
Characteristics Firmware version V1.00
CONFigure:GSM:MEAS<i>:MEValuation:SMODulation:EARea
<Status 1>, <Start 1>, <Stop 1>, <Status 2>, <Start 2>, <Stop 2>
Defines the time intervals (evaluation areas) to be used for ACP modulation
measurements.
<Status 1>, ON: Enable area 1
OFF: Disable area 1
Def. value: ON
<Start 1>, Start of evaluation area 1
Range: 0 Sym to 147 Sym
Def. value: 6 Sym
Def. unit: Sym
<Stop 1>, Stop of evaluation area 1
Range: 0 Sym to 147 Sym
Def. value: 45 Sym
Def. unit: Sym
<Status 2>, ON: Enable area 2
OFF: Disable area 2
Def. value: ON
<Start 2>, Start of evaluation area 2
Range: 0 Sym to 147 Sym
Def. value: 90 Sym
Def. unit: Sym