CMW 500 GPRF Applications
GPRF Command Reference
Operating Manual 1202.3986.32 – 03 172
7.3.3.3 Zeroing the External Sensor
REM !!! The measured RF signal must be switched off for this step !!!
FPRINT ************************************************************
FPRINT Initiate zeroing of the power sensor
FPRINT ************************************************************
CALibration:GPRF:MEASurement:EPSensor:ZERO ON
Waitkey >Sensor zeroing initiated, press OK to read sensor state
and continue<
CALibration:GPRF:MEASurement:EPSensor:ZERO?
REM !!! Switch on RF signal again to continue the measurement!!!
7.3.3.4 Continuous Sensor Measurement
FPRINT ************************************************************
FPRINT Configure a continuous measurement, return results and meas. state (should
be RUN)
FPRINT ************************************************************
INITiate:GPRF:MEASurement:EPSensor
FETCh:GPRF:MEASurement:EPSensor?
FETCh:GPRF:MEASurement:EPSensor:STATe?
7.4 GPRF Command Reference
The following sections provide detailed reference information on the
remote control commands of the General Purpose RF application. The commands are
organized as follows:
General Measurement Settings
RF Generator Commands
Power Commands
External Power Sensor Commands
For an overview of all commands refer to the Alphabetical List of Commands.
7.4.1 General Measurement Settings
The commands valid for all GPRF measurements are divided into the groups listed
below. For an overview of the commands refer to Command Groups.
Signal Routing
Analyzer Settings