EasyManua.ls Logo

Rohde & Schwarz CMW 500 - GPRF Command Reference

Rohde & Schwarz CMW 500
450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
R&S
®
CMW 500 GPRF Applications
GPRF GUI Reference
Operating Manual 1202.3986.32 – 03 152
External Attenuation (Input)
Defines the value of an external attenuation (or gain, if the value is negative) in the
input path.
The external attenuation also enters into the internal calculation of the maximum input
power that the R&S CMW 500 can measure (see "Expected Nominal Power" below).
GPIB command:
CONFigure:GPRF:MEASurement<i>:RFSettings:EATTenuation
Frequency
Center frequency of the RF analyzer. Set this frequency to the frequency of the
measured RF signal to obtain a meaningful measurement result.
GPIB command: CONFigure:GPRF:MEASurement<i>:RFSettings:FREQuency
Expected Nominal Power
Defines the nominal power of the RF signal to be measured. The nominal power is the
average output power at the DUT during the measurement intervals where the RF
transmitter is on. The "Ref. Level" is calculated as the expected peak power at the
output of the DUT:
Reference power = Expected Nominal Power + User Margin
Maximum input power
The actual input power at the connectors (i.e. the "Reference Level" minus the
"External Attenuation (Input)" value, if all power settings are configured correctly) must
be within the level range of the selected RF input connector; refer to the
"Specifications".
GPIB command: CONFigure:GPRF:MEASurement<i>:RFSettings:ENPower
User Margin
Margin that the R&S CMW 500 adds to the "Expected Nominal Power" in order to
determine its reference power; see above. The "User Margin" is typically used to
account for the known variations of the RF input signal power, e.g. the variations due
to a specific channel configuration.
The appropriate values depends on the signal properties, e.g. on the modulation
scheme.
GPIB command: CONFigure:GPRF:MEASurement<i>:RFSettings:UMARgin
7.2.2.3 Power Configuration: Measurement Control
The "Measurement Control" parameters define how the R&S CMW 500 acquires
measurement data.