R&S
®
CMW 500 WCDMA Applications
WCDMA Command Reference
Operating Manual 1202.3986.32 – 03 354
<Std Dev 1>,
...,
<Std Dev 120/240>
Magnitude error standard deviation results: 120 values for a "full
slot measurement" and 240 values for a "half slot measurement"
Range: 0 % to 50 %
D
e
f. unit: %
Example See Performing Single Shot Measurements
Characteristics Firmware version V1.00
SCPI: Query only
For additional information concerning syntax elements and returned values refer to
Multi Evaluation Measurement Commands.
9.4.2.11 Phase Error Measurement Results (Traces)
The following commands return the phase error trace results of the multi-evaluation
measurement.
FETCh:WCDMa:MEAS<i>:MEValuation:TRACe:PERRor:RMS:CURRent?
FETCh:WCDMa:MEAS<i>:MEValuation:TRACe:PERRor:RMS:AVERage?
FETCh:WCDMa:MEAS<i>:MEValuation:TRACe:PERRor:RMS:MAXimum?
READ:WCDMa:MEAS<i>:MEValuation:TRACe:PERRor:RMS:CURRent?
READ:WCDMa:MEAS<i>:MEValuation:TRACe:PERRor:RMS:AVERage?
READ:WCDMa:MEAS<i>:MEValuation:TRACe:PERRor:RMS:MAXimum?
Returns the values of the RMS phase error traces for up to 120 slots. Each value is
averaged over a half slot or a full slot, depending on the measurement period
(CONFigure:WCDMa:MEAS<i>:MEValuation:MPERiod:MODulation). The
results of the current, average and maximum traces can be retrieved.
Returned values
<Reliability>, Reliability indicator
<Phase Err 1>,
...,
<Phase Err 120/240>
RMS phase error trace results: 120 values for a "full slot
m
e
asurement" and 240 values for a "half slot measurement"
Range: -180 deg to +180 deg
Def. unit: deg
Example See Performing Single Shot Measurements
Characteristics Firmware version V1.00
SCPI: Query only
For additional information concerning syntax elements and returned values refer to
Multi Evaluation Measurement Commands.