CMW 500 GSM Applications
GSM Command Reference
Operating Manual 1202.3986.32 – 03 276
<Abs Start>, Set start level of the absolute limit for the area. The status is left
unchanged.
Range: -90 dBm to +50 dBm
Def. values: see table below
<Abs End Status> | ON: Enable the defined end level of the absolute limit
OFF: Disable the defined end level of the absolute limit
Def. values: see table below
<Abs End>, Set end level of the absolute limit for the area. The status is left
unchanged.
Range: -90 dBm to +50 dBm
Def. values: see table below
<Status Area> ON: Enable area <number>
OFF: Disable area <number>
Def. values: see table below
Characteristics Firmware version V1.00
The default settings for GSM 900/1800 are according to the following tables. The
default settings for GSM850 and GSM400 are identical to GSM900, the ones for GSM
1900 are identical to GSM 1800:
<Time
Start>
[Xs]
<Time
End>
[Xs]
<Rel Start> (=<Rel
End>) [dB]
<Abs Start> (=<Abs End>)
[dBm], <Status>
<Status
Area>
<number>=1 -38 -28 -59/-48 -36/-48, ON ON
<number>=2 -28 -18 -30 -17/-20, ON ON
<number>=3 -18 -10 -6 OFF ON
<number>=4 -10 0 4 OFF ON
Rising edge GMSK and 8PSK
<Time
Start>
[Xs]
<Time
End>
[Xs]
<Rel Start> (=<Rel
End>) [dB]
<Abs Start> (=<Abs End>)
[dBm], <Status>
<Status
Area>
<number>=1 0 542.8 1 OFF ON
<number>=2 - - - - OFF
<number>=3 - - - - OFF
Useful part GMSK
<Time
Start>
[Xs]
<Time
End>
[Xs]
<Rel Start> (=<Rel
End>) [dB]
<Abs Start> (=<Abs End>)
[dBm], <Status>
<Status
Area>
<number>=1 0 4 2.4 OFF ON
<number>=2 4 538.8 4 OFF ON