BASIC ANALYZER OPERATION
Chapter 4
............
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 123 © Veolia 2023
Figure 4-20: Injection Timing VS Integration
Methods
Each Stream and Method must be set up separately. Not all settings will be visible for
certain Methods. Some Methods are linked/or they do not have any settings as none are
needed. For example, if NPOC and TNb, all settings are tied to NPOC and TNb is not
selectable.
Use this screen to configure the “Method” settings for each stream. The Method(s) must
be enabled within the Analyzer firmware via the Service level access before it can be
configured.
Important! Sievers recommends that the Method settings are not changed after
Calibration is performed because any changes can affect results and invalidate the
Calibration.
95diag010
Injection
Injection Delay (10s)
Baseline Average
Start (10s)
Integration Delay
Start (1s)
Start Threshold Met (1ppm)
Stop Threshold
Met (1ppm)
Sample Loop
Loaded