RUNNING A GRAB SAMPLE IN OFFLINE MODE
. . . .. . . ... .... .... .... ..... ..
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 166 © Veolia 2023
Figure 4-61: Dashboard → Offline Mode → Calibration →Fit
Running a Grab Sample in Offline Mode
From the Dashboard while in Offline Mode, select the SAMPLE button. Use the following
screen to select the Outlier criteria, Repetitions, and analytical Methods prior to running
a Grab Sample measurement. When ready, select the S
TART button.
Due to the way the Analyzer sets the sample volume in the MFSV, a portion of the grab
sample may be discharged through Stream 1 or Stream 2 depending on the Analysis
Method being run for the Grab Sample. Care should be taken when running grab samples
that could directly contaminate the process sample stream.
NOTE: If only one Method has been configured in Menu → Settings →
Methods screen, that Method is used by default and the Method selection
button on this screen will not be visible.