BACKING UP THE ANALYZER SETTINGS
. . . .. . . ... .... .... .... ..... ..
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 184 © Veolia 2023
Quick Maintenance Task List
Replacing HS1, HS2, or HS3 filter or filter media
Replacing Dilution/Rinse Water
Replacing the Fan Filters
Replacing the MFSV Vent Absorber Cartridge
Replacing the Pump Tubing (GP1)
Replacing the Pump Head (GP4)
Replacing Reagent (Acid)
Replacing the Soda Lime and Activated Carbon granules (if using the Air
Scrubber)
If the maintenance task or item is not listed above as quick maintenance, consult the
“Prepare the Analyzer for Extended Maintenance” on page 196
for more information.
Backing up the Analyzer Settings
It is highly recommended to backup critical files and settings to a USB memory device
before performing any maintenance. This includes Backup Settings, Machine Report, and
taking three screenshots of different Analyzer values. This feature provides a record of
the system settings and baselines in the event there is a need to return to previous
settings.
The Analyzer must be in Offline Mode before Data Management operations are allowed.
To Backup the Analyzer Settings
1. Navigate to the following: Menu → General → Machine Configuration
2. Open the Analyzer door and insert a USB memory device into the USB Port
inside the
Analyzer
below the Touchscreen. A “U” Icon will appear in the left side panel of the
Touchcreen. This indicates a USB memory device has been connected to the
Analyzer.
3. First, select the “S” Icon on the left side panel to take a screenshot of the Machine
Configuration screen. This is Screenshot 1 of 3.
4. From the Touchscreen, select the Backup Settings S
AVE button. The Analyzer will
automatically create a Backup file of the Analyzer settings to the mounted USB
memory device.
5. Next, from the Touchscreen, select the Machine Report S
AVE button. The Analyzer
will automatically create a Backup file of the Analyzer settings to the mounted USB
memory device.
6. Navigate to the next screen: Menu → General → Software

Table of Contents

Other manuals for Veolia Sievers TOC-R3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Veolia Sievers TOC-R3 and is the answer not in the manual?

Veolia Sievers TOC-R3 Specifications

General IconGeneral
BrandVeolia
ModelSievers TOC-R3
CategoryMeasuring Instruments
LanguageEnglish