SYSTEM DESCRIPTION
Chapter 2
............
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 31 © Veolia 2023
IMPORTANT
Avoid mounting the Analyzer in direct sunlight as it may overheat the unit.
Data cabling, if required, must be installed by a qualified technician. Use of a watertight cable
bushing is required.
Avoid installing the Analyzer where it may be subject to splashing fluids. The fluids may enter the
Analyzer through the cooling system openings, causing damage.
APPLICATIONS
Wastewater
Drinking Water
Cooling Water
De-icing Water
Saline Water (Limit: 3-5% NaCl)
Process Water
Surface Water
Boiler Feed Water
Environmental Monitoring
ANALYZER HIGHLIGHTS
High-temperature combustion 1,200 °C (2192 °F)
No catalyst
No moving parts – valve system
NDIR Detection for CO
2
Multiple NDIR Detection Ranges Available for CO
2
Analytical Methods:
Included Analytical Methods: NPOC, TC
Purchasable Analytical Methods: TOCdiff / TIC, TNb, POC/VOC, TOCa
Particles: < 2 mm
Repeatability: ± 2 % End Of Detector Range
Salt Tolerance: 3-5% NaCl
Typical Cycle Time: TC ~ 1 min, NPOC ~ 3 min
1

Table of Contents

Other manuals for Veolia Sievers TOC-R3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Veolia Sievers TOC-R3 and is the answer not in the manual?

Veolia Sievers TOC-R3 Specifications

General IconGeneral
BrandVeolia
ModelSievers TOC-R3
CategoryMeasuring Instruments
LanguageEnglish