ERRORS AND WARNINGS
Appendix B
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 257 © Veolia 2023
E2364 Sampling Early Warning Stream 2 error - The result
of the TC measurement on Stream 2 will
exceed entered limit significantly
If the early warning function is
enabled; integral size 15
seconds after the crossing of
threshold exceeds the integral
size of entered limit
E2365 Sampling Check Function Stream 2 Failed
E2366 Sampling Auto Calibration Function Stream 2 Failed
Table B-5: Maintenance Request Error Codes and Descriptions
Error
Code
Module Description Criteria
E3370 Maintenance
Request
Cartridge Maintenance Required The Furnace Cartridge should
be changed at the next
possible occasion
E3371 Maintenance
Request
Pump Maintenance Required The Sample Pump's Tubing
should be changed at the next
possible occasion
E3372 Maintenance
Request
Absorber Maintenance Required The filling of the Absorbers
should be changed at the next
possible occasion
E3373 Maintenance
Request
Gas Way Maintenance Required The Gas Way should be
checked at the next possible
occasion
E3374 Maintenance
Request
Soda Lime Maintenance Required The Soda Lime should be
changed at the next possible
occasion
E3375 Maintenance
Request
Filter Mat Maintenance Required The fan's Filter Mats should be
changed at the next possible
occasion
Table B-4: Sampling Error Codes and Descriptions (Continued)
Error
Code
Module Description Criteria