TOTAL NITROGEN (TNB) PARAMETER
. . . .. . . ... .... .... .... ..... ..
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 38 © Veolia 2023
Figure 2-1: Carbon Measurement Process Diagram
Total Nitrogen (TNb) Parameter
TNb stands for Total Nitrogen Bound, which characterizes nitrogen that is attached to
inorganic and organic molecules dissolved in water. Knowledge of the amount of these
substances is important for human health and environmental reasons. Similar to TOC
analysis, the definition of dissolved TNb sets a particle size limit of 0.45 µm. Online TNb
analysis requires the same instrumentation as TOC analysis and can be performed in the
same equipment with a purchasable upgrade. The typical concentration range of interest
is between 1 and 200 mg/L TNb.
MEASUREMENT PRINCIPLE
Carbon Measurements
The basis of all Carbon measuring methods is the oxidation of the organically bound
carbon to carbon dioxide (CO
2
). This can be done thermally or through a wet chemical
process. The resulting CO
2
is detected and quantified. Thermal oxidation is ideal at
1,200 °C (2192 °F) as all substances are converted without the need for a catalyst or
chemicals. The generated CO
2
is measured in a sensitive and selective NDIR Detector.
Included Standard Operation Analytical Methods
The TC Method and NPOC Method are provided as a standard operation Analytical
Methods.
TC
Total Carbon
TIC
Total Inorganic
Carbon
TOC
Total Organic
Carbon
NPOC
Non-Purgeable Organic
Carbon
POC/VOC
Purgeable/Volatile Organic
Carbon
DOC
Non-Purgeable Dissolve
Organic Carbon
Colloids and
Particulates

Table of Contents

Other manuals for Veolia Sievers TOC-R3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Veolia Sievers TOC-R3 and is the answer not in the manual?

Veolia Sievers TOC-R3 Specifications

General IconGeneral
BrandVeolia
ModelSievers TOC-R3
CategoryMeasuring Instruments
LanguageEnglish