RUNNING A CALIBRATION IN OFFLINE MODE
. . . .. . . ... .... .... .... ..... ..
Sievers TOC-R3 Operation and Maintenance Manual
DLM 95000-01 EN Rev. A 164 © Veolia 2023
Settings
Figure 4-58: Dashboard → Offline Mode → Calibration → Settings
• Stream1 / Stream2 — Select the Stream to be Calibrated.
• Method — Select the Method(s) to be included in the Calibration.
• Add Target — add a Calibration Point. Up to four Calibration Points can be
added.
• Remove Target — reduce the number of Calibration Points, if needed.
• Repetitions — choose the number of Calibration Standard Repetitions.
• Reference — select to make the Calibration a “Reference Calibration”.
► Explanation: Marking a “Calibration as Reference”, you are marking it as
a known, good calibration. When you try to recalibrate the Analyzer, it will
compare the new calibration result to the new reference calibration result.
If the results are outside of the User-defined criteria, the Analyzer will not
allow the new calibration to be applied. For more information, see
“Calibration” on page 125.
• Start — select the START button to initiate the Calibration measurement after
confirming the parameters.