16. Next, re-tighten the Compression Nut [10A] to be finger-tight to secure the Injection
Needle in place.
17. Navigate to the Menu → Service → Sensors screen. Check the “Carrier Gas Flow
In” Sensor reading for adequate flow. For more information, see “Important Sensor
Values — Analyzer On, Analysis Stopped” on page 196.
18. When complete, “Reset” the consumable in the firmware on the Menu → Service →
Maintenance screen.
19. Once the “Cartridge” Consumable has been “Reset”, navigate to the Maintenance
(Cockpit) screen to verify that the screen displays a green, ready state for the
“Cartridge” gauge.
20. If planning to perform more maintenance, follow the procedures listed in the
appropriate sections of this chapter. Otherwise, you may turn the Furnace Power
back “On” and close the Analyzer door. The Analyzer will require at least two hours to
reach operating temperature.
Accessing the Furnace Foot Assembly
Perform this procedure if performing maintenance on or requiring access to the following
Analyzer components:
• Inspecting the Bottom of the Furnace Cartridge
• Furnace Foot Assembly