Chapter 3BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
86 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
4. Prime, as follows, using the Priming option (see Prime—Using Options Menu on page 79 for more
information):
a. Follow applicable procedure (based
on whether or not pressure sensing disc is installed) through
step to press and hold the
PRIME soft key.
b. Prime until fluid drips from end of tubing.
c. Complete procedure (installing pressure sensing disc, if applicable, and exi
ting options menu).