General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 51
Electromagnetic Compatibility
Do not use the system near magnetic resonance imaging (MRI), including Stereotaxis
technology.
Do not use the system near therapeutic radiation equipment, such as linear accelerators.
Use of any accessory, transducer or cable other than those specified can result in
inc
reased emissions or decreased system immunity and result in improper operation.
Do not use an RF device within 7.8 inches/20 cm of the radio card on the PCU. FCC
approval of the
radio card excludes co-location with any other transmitter.
Per FCC regulations, maintain a distance of at least 7.8 inches/20 cm between the radio
card on the PCU and a
human body.
Portable RF communications equipment (including peripherals such as a
ntenna cables
an
d external antennas) should be used no closer than 30 cm (12 inches) to any part of the
Alaris System, including cables. Otherwise, degradation of the performance of this
equipment could result.
The Alaris System should not be used adjacent to or stacked with other equipme
nt. If
adjacent or stacked use is necessary, monitor the Alaris System to verify that it is
operating normally in that setup.
Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the
Alaris System, including cables. Otherwise, degradation of the performance of this equipment
could result.
Interconnected data communications systems must be certified to IEC 60950 (data
processing equipment) or IEC 60601–1 (electromedical equipment).
The system is intended for use by healthcare professionals only. This is a CISPR 11 Class B
Group 1 medical system. This system can cause radio interference. Reorienting, relocating
or shielding the system, or filtering the connection to the mains network, are examples of
steps that can be taken to reduce or eliminate interference.
Medical electrical equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and used according to the EMC information
provided in Appendix A (see Compliance on page 354).
WARNINGS
CAUTIONS