Chapter 6Alaris™ EtCO
2
Module Model 8300
296 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
General Information
Warnings and Cautions
General
Microstream
TM
Disposable
WARNING
•EtCO
2
and respiratory rate readings can be affected by certain ambient environmental
and patient conditions.
The EtCO
2
module is not to be used as an apnea monitor.
The EtCO
2
module is intended only as an adjunct in patient assessment. It must be
used in conjunction with clinical signs and symptoms.
If uncertain about measurement accuracy, assess patient’s condition and vital signs
by alternate means, then ensure that the EtCO
2
module is functioning correctly.
Do not lift the EtCO
2
module by Microstream disposable because it could disconnect
from the instrument, causing it to drop on the patient. Do not place the EtCO
2
module
in any position that could cause it to fall onto the patient.
Do not use the EtCO
2
module or Microstream disposable inside a hyperbaric chamber.
Respond immediately to system alarms; patient monitoring can cease under certain
alarm conditions.
The system is not intended for use with High Frequency surgical equipment.
WARNING
Do not use a connector or Microstream disposable that appears damaged.
The Microstream disposable disconnect error message and associated alarm indicate
the Microstream disposable is disconnected. Check the Microstream disposable
connection and, if necessary, replace the Microstream disposable.
Use only Microstream disposables. Use of a disposable other than those specified can
cause improper EtCO
2
module performance, resulting in inaccurate readings.
Before use, read Microstream disposable user manual, including all warnings,
cautions, and instructions.
Carefully locate the patient Microstream disposable to reduce the possibility of patient
entanglement or strangulation.
CAUTION
Do not immerse or dampen the Microstream disposable.
The Microstream disposables are designed for single patient use and are not to be
reprocessed. Do not attempt to disinfect or flush the disposable as the EtCO
2
module can be
damaged.