10. Verify drops are flowing from the secondary container drip chamber.
NOTE:
The facility can choose to pre-populate standard drug concentrations, or leave a custom concentration
(_ _ / _ _ mL) and allow the clinician to enter the desired concentration
Patient weight or BSA is not editable during a Secondary Infusion.
Once a patient weight or BSA is entered and the infusion started, the patient weight or BSA is
automatically entered for any additional weight-based or BSA calculation. Prior to the start of infusion,
the patient weight or BSA key remains an editable field so that patient weight or BSA can be adjusted
for any module. Changing the patient weight or BSA on one module will not affect the patient weight or
BSA on any other module.
NOTE:
Programming the secondary rate over 500 mL/h
may result in c
oncurrent flow with the primary container.
NOTE:
It is possible to program an infusion with a rate that is displayed with two decimal places (one-hundredth of
a mL per hour) on the PCU for the pump module. However, due to space limitations on the pump module
rate d
isplay, the rate is displayed to the nearest one-tenth of a mL per hour on the pump module. This
value is only used for display purposes and the pump module is actually infusing at the more precise rate
noted on the PCU.
Stopping Secondary and Returning to Primary
1. Press the CHANNEL SELECT key.
2. Press the
SETUP soft key.
3. Press the
PRIMARY soft key.
4. Close clamp on secondary administration set.
or
Disconnect secondary administrati
on set fro
m upper injection port.
5. Press the
START soft key.
6. To stop secondary infusion and begin infusing primary, press the
Yes soft key.
• Secondary infusion stops and primary infusion begins.
• Main screen appears.
NOTE:
The SEC to PRI alert does not sound when the infusion is manually ended and returned to primary.