Chapter 3BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
80 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Priming Administration Set With Pressure Sensing Disc
1. Ensure that administration set is not connected to patient.
2. If installed, remove pressure sensing disc from instrument.
Using a finger
, apply firm downward pressure on pressure sensing disc (not tubing) until disc snaps
loose from slot in pressure s
ensing disc housing.
3. Press the OPTIONS key.
CAUTION
CAUTION
The pressure sensing disc, if left installed during priming, can trap air that might not be
completely expelled. To ensure that entrapped air is eliminated, it is recommended that the
pressure sensing disc be removed prior to priming and the membrane gently massaged with a
finger while priming. After priming is completed, reinstall the pressure sensing disc.
Options
Silence
Clear Cancel
Enter
System
on