Chapter 3—BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
150 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
2. To reinitiate infusion:
Press the
RESTART key.
or
Press the
CHANNEL SELECT key and then press the START soft key.
NOTE:
To stop a bolus dose, see Bolus Dose on page 98 for the procedure.
The pump module keypad is used in the illustrations but the keys are the same for the syringe module.
An infusion can also be paused by pressing the PAUSE soft key (on PCU), if the Delay Options are
disabled. To pause an infusion programmed with Delay Options enabled, see Delay Options on
page 155 and Pausing Infusion on page 158.
Changing Rate or VTBI During Infusion
1. Press the CHANNEL SELECT key.
2. Press either the
RATE or the VTBI soft key.
3. To enter desired parameter, use up/down arrows for rate titration, or numeric data entry
keys.
4. Verify correct infusion parameter entry and press the
START soft key.