Chapter 3—BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
70 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
7. Open the roller clamp.
8. Verify that no fluid is flowing through the drip chamber.
Removing
1. Close the roller clamp.
2. Open the pump module door.
The set’s safety clamp fitment automatically closes to prevent accidental free
-flow.
3. Remove set, as follows:
a. Gently pull tubing below air-in-line detector
forward and out.
b. Lift upper fitment from upper fitment receptacle.
4. If set is being removed to begin a gravity flow:
a. Depress the blue ridged release tab
on upper side of safety clamp device.
b.
Slide the white slide clamp into blue fitment (open position).
c. Adj
ust the flow rate using set’s roller clamp.
For BD Alaris pump module ensure that the tubing guide arm is not missing and
that it opens when the pump module door opens. A tubing guide arm failure can
result in a nuisance air-in-line alarm. If there is a failure, send the pump module to
CareFusion for repair.
White Slide Clamp
(shown in closed position)