General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 49
General Information
Warnings and Cautions
General
Explosion risk if used in the presence of flammable anesthetic agents or gases.
Assess patient’s condition before silencing an alarm. Do not silence alarm if patient
safety might be compromised.
Before each use, verify that the alarm limits are appropriate for the patient.
The system performs a self check during power up. The PCU should beep, no errors
should occur, and if a module is connected, all LED segments should flash. If the
system fails the self check, remove the failing PCU or module from use.
When properly secured/snapped, the release latch provides a very secure connection
between modules. If not properly latched, a module can be dislodged during operation.
Disconnect from main (AC) and battery power when performing maintenance.
Use only CareFusion batteries. The use of third party batteries could affect the safety
and efficacy of BD Alaris and Alaris products.
The battery cannot be repaired and should not be opened.
The battery is intended as a backup system. Leave the power cord connected to a
hospital grade AC power source whenever available.
Battery replacement should be performed by qualified service personnel while the
instrument is not in use.
Electrical shock hazard. Do not open case. Refer to qualified service personnel.
Due to the intermittent nature of a wireless environment, some data can be lost if a
connection cannot be established or is lost. The systems manager and wireless
network card are designed to minimize these incidents but cannot eliminate them.
The system is not intended to replace supervision by medical personnel. The user
must become thoroughly familiar with the system features, operation and accessories
prior to use.
Do not modify this equipment without authorization of the manufacturer. Authorization
should be obtained through written communication from the manufacturer.
WARNINGS