Chapter 3BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
142 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Programming — No Guardrails™ Selections
The following procedures should be used only when the drug to be infused is not listed in the drug library.
When programming a drug not listed in the drug library, the drug calculation must be programmed using
the
DRUG CALC soft key within the drug library. There are no Guardrails limits protections associated
with any non-library drug calculation.
The illustrations in this procedure assume:
ALL Mode (syringe module), Drug Calculation, Dynamic Pre
ssure Display, Profiles, and Volume
Duration configurable settings are enabled.
•NEOI
(syringe module) and Delay Options configurable settings are disabled.
Manual Programming — No Guardrails™ Selections
If Delay Options is enabled, the PAUSE soft key becomes DELAY OPTIONS.
1. Perform the following steps (see Start-Up on page 21 for specific steps):
a. Power on system.
b. Choose
Yes or No to New Patient?
c. Confirm current profile or select a new profile.
d. Enter patient identifier, if required.
2. Prepare and load syringe/administration set (see Getting started on
page 62).
3. Prime (see Getting started on page 62).
4. Start a
pplicable infusion, as described in following procedures:
No Guar
drails - Basic Infusion without Drug Calculation on
page 143
Bolus Dose on pa
ge 98