Programming
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 287
Programming
Display references throughout this procedure are for illustration purposes only.
See General Information on page 49,
Features and Displays on page 52, and BD Alaris™ PCU Model
8015 and Alaris™ PCU Model 8015 on page 15 fo
r information about:
•Displays
Operating Features, Controls, Indicators
Monitoring Mode
1. Perform the following steps (see BD Alaris™ PCU Model 8015 and Alaris™ PCU Model 8015 on
page 15, General Setup and Operation on page 18, and Sta
rt-Up on page 21 for more information):
a. Power on system.
b. Choose
Yes or No to New Patient?.
c. Confirm current profile or select a new profile.
d. Enter patient identifier, if required.
2. Connect Microstream disposable (see Getting Started on pa
ge 284).
3. Press
CHANNEL SELECT key.
SENSOR WARMING and then SEARCHING appear in Channel Message display until EtCO
2
and
respiratory rate readings stabilize (up to 60 seconds).
A
L
A
R
M
M
O
NIT
O
R
S
T
A
N
D
B
Y
CHANNEL
SELECT
MONITOR
CHANNEL
OFF
EtCO (mm Hg)
2
RR (breaths/min)