General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 279
Specifications and Symbols
Specifications
Models 8210 and 8220
Alarms:
Audible and visual alarms for high and low saturation and pulse rate, sensor condition,
system failure, and low battery conditions.
Alarm Limits:
Dimensions:
3.3" W
x
8.9" H
x
5.5" D
(
8.4
cm
W
x
22.6
cm
H
x
14
cm
D
)
Environmental Conditions:
Fluid Ingress Protection:
IPX1
, The SpO
2
module is protected against harmful effects of dripping water per
IEC 60529
Mode of Operation:
Continuous
Shock Protection:
Type BF patient applied part
Weight:
2.5
lbs (
1.13
kg)
Low High
Pulse Rate: 30 - 239 bpm 31 - 240 bpm
SpO
2
20 - 99% 21 - 100%
Symbols Meaning Operating
Storage/
Transport
Atmospheric
Pressure
525 - 4560 mmHg
(70 - 608 kPa)
375 - 760 mmHg
(50 - 101 kPa)
Relative Humidity
(Avoid prolonged
exposure to relative
humidity >90%)
20 - 90%
Noncondensing
5 - 90%
Noncondensing
Temperature Range 41 - 104°F
(5 - 40°C)
-4 - 140°F
(-20 - 60°C)