Appendix ATroubleshooting and Maintenance
346 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Cleaning
Inspect and clean the product per the following procedures. Read all warnings and cautions before
continuing with this procedure.
BD Alaris™ and Alaris™ System
Cleaning Products
Refer to the following website for cleaning product information:
www.carefusion.com/alarissystemcleaning.
Procedure
Inspect IUI Connectors
WARNING
To prevent an electrical hazard:
Turn the instrument off and unplug the power cord from AC power before cleaning.
Do not spray fluids directly onto the instrument or into the IUI connectors.
D
o not steam autoclave, EtO sterilize, immerse the instrument in fluids, or allow
flui
ds to enter the instrument case.
Do not connect a module until the IUI connectors are thoroughly dry
.
Do not use compressed air to dry the instrument; this could force fluid into the
instrument.
WARNING
Inserting a finger or other object into the IUI connector, when the module is attached to the PC
unit, could result in electrical shock.
WARNING
Failure to perform these inspections can result in improper instrument operation.
CAUTION
CAUTION
Preventive maintenance inspections should only be performed by qualified service personnel.