Chapter 3—BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
186 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
• Total dose:
Drug amount.
Drug amount / patient body surface area (
BSA).
Drug amount / patient weight.
Configurable Settings
See BD Alaris™ PCU Model 8015 and Alaris™ PCU Model 8015 on page 15 for system configurable
settings.
If the configuration settings need to be changed from the Factory default settings, refer to the appl
icable
Technical Service Manual or contact CareFusion technical support, for technical, troubleshooting, and
preventive maintenance information.
With the profiles feature enabled, the settings are configured independently for each profile. A
hospital-defined, best- practice data set must be uploaded to enable the profiles feature. Date and Time is a
system setting and is the same in all profiles.
Shared Infusion
x
Pump Module
Feature Default Setting Options
Delay Options
• Callback
Disabled
After
Enabled - Disabled
None, Before, After, Before and After
Drug Calculation
• Bolus Dose
Disabled
Disabled
Enabled - Disabled
Enabled - Disabled
Pressure Dynamic
(Dynamic Pressure Display)
Disabled Enabled - Disabled
Volume/Duration Disabled Enabled - Disabled
Feature Default Setting Options
Accumulated Air Enabled Enabled - Disabled
Air-in-Line Settings
(single bolus)
75 mcL 50, 75, 125, 175, or 250 mcL
Anesthesia Mode only: 500 mcL
Auto-Restart Attempts 0 0 - 9 attempts
Anesthesia Mode only: 9 attempts
KVO (Keep Vein Open) 1 mL/h
0.1
-
20
mL/h
Max Rate 999 mL/h
0.1
-
99.9
mL/h in
0.1
mL/h increments;
100
-
999
mL/h in
1
mL/h increments
Max VTBI 9999 mL
0.1
-
9999
mL