General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 197
Graphs
Pump Module
NOTE:
The plot range has been increased to ±100% to allow visualizati
on of the graph.
-0.10
-0.05
0.00
0.05
0.10
0.15
0.20
0.25
0.30
0 20 40 60 80 100 120
Flow Rate (mL/h)
Time (min)
Start-up at 0.1 mL/h (iniƟal)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
2.0
0 20 40 60 80 100 120
Flow Rate (mL/h)
Time (min)
Start-up at 1 mL/h (ŝŶŝƟĂůͿ
-80.7
-61.6
-44.9
-27.4
-25.9
62.3
38.2
11.6
-1.6
-6.8
-14.2
-120.0
-80.0
-40.0
0.0
40.0
80.0
120.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽŶ /Ŷƚerval (miŶͿ
Trumpet Curve at 0.1 mL/h (iŶŝƟĂůͿ
-36.6
-16.8
-6.3
-5.6
-3.1
35.5
13.0
2.6
2.1
-0.1
-2.5
-45.0
-30.0
-15.0
0.0
15.0
30.0
45.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽn Interval (min)
Trumpet Curve at 1.0 mL/h (inŝƟĂůͿ
-101.3
-81.8
-41.8
-31.6
-18.7
85.9
56.2
28.3
15.7
3.6
-6.2
-120.0
-80.0
-40.0
0.0
40.0
80.0
120.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽŶ /Ŷƚerval (miŶͿ
Trumpet Curve at 0.1 mL/h (96 hrs)
-28.7
-14.5
-5.2
-5.1
-2.4
31.3
11.9
1.5
1.7
-0.3
-1.6
-45.0
-30.0
-15.0
0.0
15.0
30.0
45.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽn Interval (min)
Trumpet Curve at 1.0 mL/h (96 hrs)
Legend:
Maximum rate error
Overall rate error
Minimum rate error