Programming
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 215
10. Verify correct parameters on second nurse summary screen and press START soft key.
Infusion mode and PCA drug name scroll in Channel Messa
ge Display. If a loading dose has been
entered, scrolls
DELIVERING LOAD.
Main Display alternates between volume remaining and PCA drug name with infusion mode.
When PCA dose is delivered:
°
Green Infusing Status Indicator illuminates.
°
Rate display flashes "_ _ _ _ _".
°
DELIVERING PCA scrolls in Channel Message Display.
°
When PCA dose is complete, PCA COMPLETE scrolls in Channel Message Display.
Setting Up Continuous Infusion Only
1. Perform steps in Preparing Infusion on page 208.
2. Press
CONTINUOUS INFUSION soft key from Infusion Mode screen.
[Conc]: 1 mg/mL
Second Nurse Summary
Morphine
A
>Press START
START
PAUSE
PRO-
GRAM
10
minutes
Lockout
Interval:
Max Limit:
2 mg
PCA Dose:
25
mg/4 h
PCA Dose only
Loading Dose Only
A
Continuous Infusion
>Select an Option
Guardrails Drug Setup
Morphine
PCA
Dose + Continuous
INFUSION MODES
SETUP
DRUG
LIBRARY

Table of Contents