Chapter 4—Alaris™ PCA Module Model 8120
216 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
3. To enter Continuous Infusion dose, press CONT DOSE soft key and use numeric data entry keys.
4. To enter maximum limit, press MAX LIMIT soft key, press Yes soft key, and use numeric data entry keys.
Time (in hours) associated with
Max Limit is automatically entered based on setup in system
configuration.
If
No is selected, no MAX LIMITS are available for this infusion.
5. To enter loading dose, press LOAD DOSE soft key, press Yes soft key, and use numeric data entry keys.
Loading dose is included in volume infused but is not included in
Max Limit.
A
>Enter Continuous Dose
MAX
LIMIT
_ _ _ mg/h
CONT
DOSE
[Conc]: 1 mg/mL
LOADING
DOSE
None
CHANGE
MODE
Continuous
Morphine
None
A
[Conc]: 1 mg/mL
Continuous
Morphine
>Select Max Limit Option
2 mg/h
MAX
LIMIT
CONT
DOSE
LOADING
DOSE
None
CHANGE
MODE
> > > >
Yes
No