Programming
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 217
6. Verify correct parameters and press CONFIRM soft key.
• If the programmed parameters are outside the soft limit for that c
are area, an audio alert sounds
and a visual prompt appears before programming can continue. If
Yes soft key is pressed,
programming continues; if
No soft key is pressed, infusion needs to be reprogrammed.
• If the programmed parameters are outside the hard limit for that care area
, an audio alert sounds
and a visual prompt appears before programming can continue. Infusion must be reprogrammed.
• If there is a potential for a programmed "_ _ / _ _ mL" parameter to result in an excessiv
e volume
or dose being delivered, following prompt appears:
Cannot proceed due to incorrect concentration or dosing parameters. Remove syringe, verify
concentration, and reprogram.
The prompt can be (a) the result of an incorrect drug amount and/or diluent volume entry, or
(b) can occur if hospital-established Guardrails limits are very wi
de. Be sure to enter either a drug
amount per
1 mL or total drug amount per total volume—for example, a 30 mL syringe with a
concentration of 1 mg/1 mL can be entered in one of two ways:
Drug Amount 1 mg
Diluent Volume 1 mL
or
Drug Amount 30 mg
Diluent Volume 30 mL
• If a soft limit is overridden,
G icon is displayed. When G soft key is pressed, all applicable out-of-
range limits are listed.
7. Close and lock security door.
8. Verify correct programming pa
rameters and press
START soft key.
• Green Infusing Status Indicator illuminates.
• Infusion mode and drug name scroll in Channel Message Display
. If a loading dose has been
entered,
DELIVERING LOAD scrolls.
• Volume infused in mL/h is displayed in Rate Display.
• Main Display alternates between volume remaining
and infusion mode with drug name.
A
[Conc]: 1 mg/mL
>Press START
Second Nurse Summary
Morphine
1 mg
START
PAUSE
PRO-
GRAM
Loading
Dose:
Max Limit:
2 mg/h
Cont. Dose:
30
mg/4 h