Programming
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 217
6. Verify correct parameters and press CONFIRM soft key.
If the programmed parameters are outside the soft limit for that c
are area, an audio alert sounds
and a visual prompt appears before programming can continue. If
Yes soft key is pressed,
programming continues; if
No soft key is pressed, infusion needs to be reprogrammed.
If the programmed parameters are outside the hard limit for that care area
, an audio alert sounds
and a visual prompt appears before programming can continue. Infusion must be reprogrammed.
If there is a potential for a programmed "_ _ / _ _ mL" parameter to result in an excessiv
e volume
or dose being delivered, following prompt appears:
Cannot proceed due to incorrect concentration or dosing parameters. Remove syringe, verify
concentration, and reprogram.
The prompt can be (a) the result of an incorrect drug amount and/or diluent volume entry, or
(b) can occur if hospital-established Guardrails limits are very wi
de. Be sure to enter either a drug
amount per
1 mL or total drug amount per total volume—for example, a 30 mL syringe with a
concentration of 1 mg/1 mL can be entered in one of two ways:
Drug Amount 1 mg
Diluent Volume 1 mL
or
Drug Amount 30 mg
Diluent Volume 30 mL
If a soft limit is overridden,
G icon is displayed. When G soft key is pressed, all applicable out-of-
range limits are listed.
7. Close and lock security door.
8. Verify correct programming pa
rameters and press
START soft key.
Green Infusing Status Indicator illuminates.
Infusion mode and drug name scroll in Channel Message Display
. If a loading dose has been
entered,
DELIVERING LOAD scrolls.
Volume infused in mL/h is displayed in Rate Display.
Main Display alternates between volume remaining
and infusion mode with drug name.
A
[Conc]: 1 mg/mL
>Press START
Second Nurse Summary
Morphine
1 mg
START
PAUSE
PRO-
GRAM
Loading
Dose:
Max Limit:
2 mg/h
Cont. Dose:
30
mg/4 h