Programming
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 289
5. To confirm alarm settings and return to ETCO2 Main display, press CONFIRM soft key.
6. To return to Main Display, press MAIN SCREEN soft key.
Navigating Trend Data
1. To view Trend Data, press TREND soft key.
.
The following information is displayed:
• TIME period for data review.
•Av
erage ETCO2 with high and low values.
• A
verage respiratory rate (RR) with high and low valu
es.
•Alarm icon () with Fi in TIME column to
indicate high FiCO
2
alarm limit has been exceeded.
•Alarm icon () to indicate an alarm limit has been exceeded.
•Al
arm icon () in RR column to indicate a no breath ( ) alarm limit has been triggered.
• D
ashes (---), if no EtCO
2
or respiratory rate values are available for time period displayed.
Tabular information is not
updated while Trend Data view is displayed. Tabular data is updated, using
new trend data stored in the EtCO
2
module, after leaving Trend Data view. To view latest data, return
to
Trend Data view.
2. To navigate from page to page, press
PAGE UP and PAGE DOWN soft keys.
A
ETCO2
LOW
35
_40
CONFIRM
Adult
Mode
Alarm Limits
25
5
RR
HIGH
RR
LOW
10
NO
BREATH
mmHg
mmHg
breaths/min
breaths/min
seconds
FICO2
LIMIT
FiCO2 High : 10 mmHg
ETCO2
HIGH
>Press CONFIRM to
Apply Changes
A
>Press UP/DOWN Keys
to Move Cursor.
PAGE
DOWN
ZOOM
ETCO2
MAIN
ZOOM: 60 5 minutes
TIME
RRETCO2
AVG
MIN
MAX
AVG
MIN
MAX
2001-07-06
21:28
20:58
20:28
22:58
40
38
47
41
34
45
---
39
37
44
13
11
26
11
10
14
---
12
11
14
22:28
40
44
39
12
16
21:58
41
39
45
13
11
20
1120
Trend Data
Adult Mode
09:00
30
PAGE
UP
Fi
Fi