Chapter 4Alaris™ PCA Module Model 8120
244 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Administration Set/Syringe Information
The PCA module uses standard, single-use, disposable syringes (with luer-lock connectors) and
administration sets with anti-siphon valves, designed for use on syringe-type PCA module pumps.
For specific administration se
t instruct
ions and set
replacement interval, refer to user manual provided
with set.
For a list of compatible syringes, refer to Compatible Syringes on page 244.
Use aseptic techniques when handling sets and syringes.
Administration sets are supplied with a sterile a
nd nonpyrogenic fluid path for one-time use. Do not
resterilize.
Discard administration set per facility protocol.
•For
IV push medication (put instrument on hold), clamp tubing above port.
Flush port(s) per facility protocol.
Compatible Syringes
The PCA module is calibrated and labeled for use with the following single-use disposable luer-lock
syringes. Use only the syringe size and type specified on the Main Display. The full list of permitted
syringe models is dependent on the PCA module’s software version.
Syringe variability can impact occlusion pressure sensing.
The variability can reduce the device’s time to
alarm and/or might req
uire that a higher alarm pressure limit be programmed.
NOTE:
Prefilled morphine sulfate 1 mg/mL.
Manufacturer 20 mL 30 mL 35 mL 50 mL 60 mL
BD Plastipak
xx xx
IMS Pump Jet
x
Monoject
xxx
Terumo
xx x