General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 187
Syringe Module
NOTE:
*Occlusion Pressure Point cannot be set higher
than the default setting.
Pressure Mode
Mode Selection
L
ock Status
Max Occlusion Pressure
Default Starting Occlusion
Pressure
Pump
Unlocked
525 mmHg
525 mmHg
Pump, Selectable
Locked, Unlocked
50 - 525 mmHg in 25 mmHg increments (adjustable only
in
Selectab
le Pressure Mode)
50 - 525 mmHg in 25 mmHg increments (configured by
profile and ad
justable only in Selectable Pressure Mode)
SEC to PRI Alert Enabled Enabled - Disabled
Secondary (Dual Rate Sequential
Piggybac
king)
Di
sabled Enabled - Disabled
Feature Default Setting Options
ALL Mode Disabled Enabled - Disabled
Auto Pressure Disabled Enabled - Disabled
Back Off (after occlusion) Enabled Enabled - Disabled
Fast Start Enabled Enabled - Disabled
KVO (Keep Vein Open)
Rate Adjust
Volume Adjust
Disabled
1 mL/h
5%
Enabled - Disabled
0.01 - 2.5 mL/h (0.01 - 0.09 mL/h available for 1 mL and
3 mL syringes)
0.5 - 5% (Based on the total syringe size, not the volume of
fluid in
the syringe)
Max Rate 999 mL/h 0.1 - 99.9 mL/h in 0.1 mL/h increments;
100 - 999 mL/h in 1 mL/h increments
Near End (NEOI)
Continuous Alert Time
I
ntermittent NEOI
Intermittent Alert Time
Disabled
60
Disabled
15
Enabled - Disabled
1 - 60 minutes
Enabled - Disabled
1 - 60 minutes
NEOI Snooze
(Shared syringe and PCA setting)
Disabled Enabled - Disabled
5, 10, or 15 minutes
Occlusion Pressure Set Point
With Disc
N
o Disc
1000 mmHg
High*
25
-
1000
mmHg in
1
mmHg increments
Low (200 mmHg)
Medium (500 mmHg)
High (800 mmHg)
Priming Disabled Enabled - Disabled
Feature Default Setting Options