Chapter 3BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
188 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Specifications
Pump Module
Accumulated Air
Window:
NOTE:
In a standard bore tubing set:
50 mcL is approximately 0.34
inches of air
75 mcL is approximately 0.51
inches of air
125 mcL is approximately 0.85 inc
hes of air
175 mcL is approximately 1.19 inc
hes of air
250 mcL is approximately 1.7 inches o
f air
In Anesthesia Mode only.
Bolus Volume, Maximum
after
Occlusion:
Critical Volume:
Maximum over-infusion that can occur in the event of a single fault condition is 0.6 mL.
Dimensions:
3" W x 8.8" H x 5.5" D
Environmental
Conditions:
Single Bolus
Setting
(mcL)
Volume
Window (mL)
% Air that
Causes Alarm
Volume (mL) that
Causes Alarm
50 2.8 10% 0.28
75 8.0 20% 1.6
125 8.0 20% 1.6
175 8.0 20% 1.6
250 8.0 30% 2.4
500
12.0 30% 3.6
Pressure Limit (mmHg) Rate (mL/h) Bolus Volume (mL)
50 25
0.3
525 25
0.6
Symbols Meaning Operating Storage/Transport
Atmospheric
Pressure
525 - 4560 mmHg
(70 - 608 kPa)
375 - 760 mmHg
(50 - 101 kPa)
Relative Humidity
(Avoid prolonged
exposure to relative
humidity >90%)
20 - 90%
Noncondensing
5 - 90%
Noncondensing
Temperature Range 41 - 104°F
(5 - 40°C)
-4 - 140°F
(-20 - 60°C)