4. To confirm profile selection, press the CONFIRM soft key.
Main screen appears.
Updating the Data Set
It is important to ensure that the latest data set is loaded into the PCU. To check the status of the data set,
see Viewing Data Set Status on pa
ge 47. If the data set status is Pending, use the following steps to update
the data set.
1. Power off the PCU (also referred to as power
-cycling).
a. Press the
OPTIONS key.
b. Press the
Power Down All Channels soft key.
c. Press
Yes at the confirmation prompt.
2. Power on the PCU.
3. Press the
Yes soft key at the New Patient screen. Doing this clears all stored patient parameters from the
memory and activates the pending data set.
4. Review the data set status. It should now display
Current.
5. To confirm data set status:
a. Press the
OPTIONS key.
b. Press the
Page Down soft key, and select Data Set Status.
Data set status should display
Current.
If data set status is
Pending, repeat step 1.
Midtown Hospital
Profiles
>Press CONFIRM
CONFIRM
PAGE
DOWN
Neonatal
Peds ICU
Neonatal ICU
Adult General Care
Adult ICU
1 of 2