Chapter 3BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
62 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
Getting started
Introduction
This section of the user manual provides pump module (model 8100) and syringe module (model 8110)
instructions and information. It is used in conjunction with:
BD Alaris™ administration set instructions
Drug product labeling
PCU chapter of this manual (see Getting started o
n pa
ge 62)
Alaris™ Pump Module Set Compatibility Car
d
Alaris™ Syringe Module Set Compatibility Car
d
BD Alaris™ PC Unit Model 8015, Alaris™ PC Unit Model 8015, BD Al
aris™ Pump Module Model
8100, and Alaris™ Pump Module Model 8100 Technical Service Manual
Syringe Module Model 8110 and PCA Module
Model 8120 Technical Service Manual
System maintenance software (and its instructions) for system che
ck-in, maintenance, and
configurations for connecting the PCUs to the wireless network
Essential Performance
The modules are designed to accurately deliver the programmed amount of the medication or fluid over the
programmed time period. The modules ensure that an infusion is not being inadvertently delivered when
the user expects the system to be in a paused, stopped, or off condition. The modules employ measurement
systems to detect and alarm for conditions adverse to safe administration of fluid. These include
measurements of proper infusion set loading (free flow detection), pressure (occlusion detection), and
air-in-line detection.
The EtCO
2
and SpO
2
modules monitor and alarm when a patient physiological condition is out of range.
The pump and syringe modules are intended for
facilities that utilize infusion and/or syringe pumps for the
delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery
through clinically acceptable routes of administration, such as - intravenous (IV), intra-arterial (IA),
subcutaneous, epidural, enteral, or irrigation of fluid spaces. The pump and syringe modules are indicated
for use on adults, pediatrics, and neonates. Up to four pump and/or syringe modules can be connected to
the system.
Administration Sets/Syringes: See Administration Set/Syringe
Information on page 173 for specific
administration set and syringe information.
Alarms and Alerts: See Alarms and Alerts on
page 325
(Appendix A) for module-specific alarms and
alerts.
Contraindications: None kn
own.
WARNING
Read all instructions, for both the infusion modules and PCU, before using the system.