EasyManua.ls Logo

BD Alaris - Page 228

Default Icon
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4Alaris™ PCA Module Model 8120
214 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
8. Verify correct parameters and press CONFIRM soft key.
If the programmed parameters are outside the soft limit, an audio al
ert sounds and a visual prompt
appears before programming can continue. If
Yes soft key is pressed, programming continues; if
the
No soft key is pressed, infusion must be reprogrammed.
If the programmed parameters are outside the hard limit for that care area
, an audio alert sounds
and a visual prompt appears before programming can continue. Infusion must be reprogrammed.
If there is a potential for a programmed “_ _ / _ _ mL” parameter to result in an excessive volume
or dose
being delivered, following prompt appears:
Cannot proceed due to incorrect concentration or dosing parameters. Remove syringe, verify
concentration, and reprogram.
The prompt can be (a) the result of an incorrect drug amount and/or diluent volume entry, or
(b) can occur if hospital-established Guardrails limits are very wide. Be sure to enter either a
drug amount per 1 mL or total drug amount per total volume—for example, a 30 mL syringe
with a concentration of 1 mg/1 mL can be entered in one of two ways:
Drug Amount 1 mg
Diluent Volume 1 mL
or
Drug Amount 30 mg
Diluent Volume 30 mL
If a soft limit is overridden,
G icon is displayed. When G soft key is pressed, all applicable out-of-
range limits are listed.
9. Close and lock security door.
[Conc]: 1 mg/mL
_ 25 mg/4 h
10 minutes
Morphine
PCA Only
LOCKOUT
INTERVAL
MAX
LIMIT
A
>Close and Lock Door Then
Press CONFIRM
2 mg
CONFIRM
PCA
DOSE
LOADING
DOSE
None
CHANGE
MODE

Table of Contents

Other manuals for BD Alaris

Related product manuals