EasyManua.ls Logo

BD Alaris - Page 212

Default Icon
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3—BD Alaris™ Pump Module Model 8100, Alaris™ Pump Module Model 8100, and Alaris™
Syringe Module Model 8110
198 BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual
0
5
10
15
20
25
30
35
40
45
50
0 20 40 60 80 100 120
Flow Rate (mL/h)
Time (min)
Start-up at 25 mL/h (ŝŶŝƟĂůͿ
0
200
400
600
800
1000
1200
1400
1600
1800
2000
0 20 40 60 80 100 120
Flow Rate (mL/h)
Time (min)
Start-up at 999 mL/h (iniƟal)
-2.6
-1.8
-1.4
-1.3
-1.2
-0.1
-0.6
-0.9
-1.0
-1.0
-1.1
-15.0
-10.0
-5.0
0.0
5.0
10.0
15.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽn Interval (min)
Trumpet Curve at 25 mL/h (inŝƟĂůͿ
-2.0
-1.8
-1.7
-1.6
-1.6
-1.2
-1.4
-1.5
-1.5
-1.5
-1.6
-15.0
-10.0
-5.0
0.0
5.0
10.0
15.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽn Interval (min)
Trumpet Curve at 999 mL/h (inŝƟĂůͿ
-3.5
-2.2
-1.6
-1.3
-1.1
0.2
-0.2
-0.6
-0.7
-0.8
-0.9
-15.0
-10.0
-5.0
0.0
5.0
10.0
15.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽn Interval (min)
Trumpet Curve at 25 mL/h (96 hrs)
-2.3
-1.9
-1.7
-1.6
-1.5
-0.9
-1.2
-1.3
-1.4
-1.4
-1.5
-15.0
-10.0
-5.0
0.0
5.0
10.0
15.0
0 5 10 15 20 25 30
Flow Rate Error (%)
ObservĂƟŽn Interval (min)
Trumpet Curve at 999 mL/h (24 hrs)
Legend:
Maximum rate error
Overall rate error
Minimum rate error

Table of Contents