EasyManua.ls Logo

BD Alaris - Page 213

Default Icon
376 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Information
BD Alaris
TM
System with Guardrails
TM
Suite MX User Manual 199
Syringe Module
-5.3
-4.0
-3.0
-2.6
-2.5
1.5
-0.1
-1.1
-1.7
-1.8
-2.1
-15
-10
-5
0
5
10
15
0 5 10 15 20 25 30
Flow Rate Error (%)
Observation Interval ( min )
Trumpet Curve at 1 mL/h (Initial)
-1
0
1
2
3
4
5
6
7
8
9
10
0 20406080100120
Flow Rate ( mL / h )
Start-Up Curve at 5 mL/h (Initial)
-5.3
-4.0
-3.0
-2.6
-2.5
1.5
-0.1
-1.1
-1.7
-1.8
-2.1
-15
-10
-5
0
5
10
15
0 5 10 15 20 25 30
Flow Rate Error (%)
Observation Interval ( min )
Trumpet Curve at 1 mL/h (Initial)
-4.2
-2.1
-1.9
-1.6
-1.3
3.5
0.7
0.1
-0.4
-0.7
-1.1
-15
-10
-5
0
5
10
15
0 5 10 15 20 25 30
Flow Rate Error (%)
Observation Interval ( min )
Trumpet Curve at 5 mL/h (Initial)
Legend:
Maximum rate error
Overall rate error
Minimum rate error

Table of Contents